GDKS-205A Automatic Flash Point Open Cup Tester
● 5-inch TFT LCD touch screen.
● With high speed signal processor control, reliable and high accuracy.
● Detection, ignition, alarm, cooling, printing, the whole testing process is completed automatically.
● Platinum electric wire or gas ignition.
● Built-in atmosphere pressure sensor, auto-measure and calibrate result.
● Self-adaptive PID control calculation, automatic adjusting temperature rise curve according to standard requirements.
● Built-in multiple implementation standards such as GB/T3536-2008, ASTM D92 and custom, and users can choose according to their needs.
● With over-temperature protection function. When the set temperature is exceeded, it will automatically stop heating.
● Built-in high speed thermal printer.
● Store up to 500pcs history records.
● After the test, the instrument can automatically start the air cooling function, and the cooling speed is fast.
Temperature Control Range: | l Room Temperature--400℃ |
Repeatability: | No More Than±3℃ |
Resolution: | 0.1℃ |
Accuracy: | 0.5% |
Power Voltage: | AC220V±20% 50Hz±10% |
Max.Power: | <400W |
Test Standards: | GB/T3536-2008, ASTM D92 |
Environmental Temperature: | 5℃-40℃ |
Relative Humidity: | ≤85% RH |
Ignition Method: | Gas Flame Or Electric Ignition |
Weight: | 10kg |
Dimension: | 440*280*290mm |
Tester | 1 set |
Oil cup | 1 pc |
Fuse | 2 pcs |
Power cord | 1 pc |
Print paper | 1 roll |
User’s guide | 1 copy |
Packing list | 1 copy |
Test report | 1 copy |